Instruction Manual

6 ZOLL M Series Temperature
Specifications
Table 2: General Specifications
Principle of Operation Thermistor - YSI 400 compatible only
Table 3: Temperature Measurement Specifications
Temperature Measurement Range 59° to 113° F (15° to 45° C)
Accuracy +/- 0.09° F (0.05° C) (not including sensor)
Sampling Rate 1 sample/sec (1 Hz)
Resolution 0.1° F (0.06° C)
Table 4: Battery Operating Time Specifications (using ZOLL XL Battery Pack)
For a new, fully charged battery pack at
20°C with IBP, EtCO
2
, SpO
2,
NIBP and
Temperature options
60 defibrillator discharges at maximum energy (200J
Biphasic), or 1.75 hours minimum of continuous ECG,
IBP (2 channels), EtCO
2
, SpO
2
, and Temperature
monitoring, or 2.00 hours of continuous ECG
monitoring/pacing at 60 mA, 70 beats/min.
Table 5: Environmental Specifications
Operating Temperature 32° to 122° F (0° to 50° C)
Storage and Shipping Temperature -4° to 140° F (-20° to 60° C)
Electromagnetic Immunity AAMI DF-2: 1996, EN 61000-4-3: 1996, 15 V/m
Humidity 5 to 95% relative humidity, non-condensing
Operating Pressure 594 to 1060 mBar
Table 6: Alarm Limits Specifications
Temperature High 59.2 to 113.0° F
Low 59.0 to 112.8° F
High 15.1 to 45.0° C;
Low 15.0 to 44.9° C
Temp (difference in temperature)
High 0.2 to 54.0° F
Low 0.0 to 53.8° F
High 0.1 to 30.0° C;
Low 0.0 to 29.9° C