Configuration Option Tables
31
Advisory and Semi-automatic Modes
Feature Options Default
Energy Level: Shock 1 Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J
Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J
Manual DSW: 200 J
Manual Biphasic: 120 J
AED DSW: 200 J
AED Biphasic: 120 J
Energy Level: Shock 2 Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J
Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J
Manual DSW: 300 J
Manual Biphasic: 150 J
AED DSW: 200 J
AED Biphasic: 120 J
Energy Level: Shock 3 Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J
Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J
Manual DSW: 360 J
Manual Biphasic: 200 J
AED DSW: 360 J
AED Biphasic: 200 J
Auto Analyze 3 Times Yes, No, Only in Semi-Auto Manual: No
AED: Only in Semi-Auto
“NO SHOCK ADVISED” Prompt “NO SHOCK ADVISED”, “CHECK PULSE”,
BOTH
“NO SHOCK ADVISED”
“CHECK PATIENT” Prompt “CHECK PATIENT”, “PRESS ANALYZE” “CHECK PATIENT”
Display “DO CPR” Message Yes, No, Only in Semi-Auto Manual: No
AED: Only in Semi-Auto
Display ECG in Semi-Auto Mode Yes/No Yes
Store to Card in Auto Monitor Mode Yes/No Yes
Auto Charge in Advisory Mode Yes/No Yes
Record to Card in Advisory Mode None, ECG Only, Voice and ECG None
Enable Voice in Advisory Mode Yes/No No
Display Heart Rate in Semi-Auto Mode Yes/No No
Auto-Analyze on Power Up Yes/No No
Enable 12-Lead in Semi-Auto Mode Yes/No Yes
AED Defib Ready Hold Time 15 seconds, 60 seconds 15 seconds