User Manual

Configuration Option Tables
29
Configuration Option Tables
The following tables list the M Series CCT options and their possible values.
An asterisk (“*”) refers to features that are currently not implemented on the M Series CCT. These features will only
operate in accordance with their pre-configured default settings (listed below).
General
Feature Options Default
Selected Language English, Other English
Notch Filter (ECG) 50, 60 Hz 60 Hz
Allow Card Erase Yes/No Yes
Display Elapsed Time Yes/No AED: Yes
Manual Advisory: No
* Voice Markers Enabled Yes/No No
Set Report Restart Delays Set/Review Set/Review
Summary Report Restart Delay 5, 15, 30, 90 minutes, 1.5 days 15 minutes
Trend Report Restart Delay 5, 15, 30, 90 minutes, 1.5 days 15 minutes
Print 12 Lead 4x3 None, 1 Copy, 2 Copies None
* Auto Self-Test TIme-out Period No Test, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17,
18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29,
30 days
No test
Async Softkey in Pace Mode Yes/No Yes
Code Markers
(Defib Code Markers, Monitor Code
Markers, Pace Code Markers)
Config/Review Config/Review
Device Identifier 0 through 9, A through Z, blank, and “_” inclusive
for each digit
0 0 0 0 0 0 0 0 0 0 0
Manual Mode Passcode 0 through 3 inclusive, for each digit 0 0 0
Configuration Mode Passcode 0 through 3 inclusive, for each digit 0 0 0 0 0 0 0 0
Display Lead on Power-Up Paddles/Pads, I, II, or III Paddles/Pads
Frequency Response 0.5 - 21 Hz, 0.5 - 27 Hz, 1 - 21 Hz 0.5 - 21 Hz
QRS Volume on Power-Up Midrange, OFF Midrange
Pace Rate Setting on Power-Up 30 through 102 ppm (increments of 4 ppm) 70 ppm
Displayed Implanted Pace Spikes Yes/No Yes
Gain on Power-Up (ECG) AGC, 0.5 cm/mV, 1 cm/mV, 1.5 cm/mV, 2 cm/mV AGC
Report Memory Card Errors Yes/No Yes
Enable Leads Off Yes/No Yes
Upload Baud Rate 9600 bps, 38400 bps, 115200 bps 38400 bps
Trending Report Zoom Level All, 5 min., 10 min., 15 min., 30 min., 60 min. All
Trending Enabled Yes/No Yes
Battery Type Used PD 4410, XL Battery PD 4410
Alarm Limits at Power-Up Set/Review Set/Review
Alarms Active at Power-Up Yes/No No
Low Battery Tone 1, 5 minutes 1 minute