CONFIGURATION GUIDE 9650-0204-01 Rev A October 2006
An issue or revision date for this manual is shown on the front cover. If more than three years have elapsed since this date, contact ZOLL Medical Corporation to determine if additional product information updates are available. ZOLL, M Series, and stat.padz are registered trademarks, and PD, and XL are trademarks of ZOLL Medical Corporation. 12SL and Catalyst MUSE are trademarks of GE Medical Systems. Copyright 2006 by ZOLL Medical Corporation. All rights reserved.
Table of Contents INTRODUCTION............................................................................................................................. 1 CONFIGURING THE UNIT ............................................................................................................1 SELECT LANGUAGE 2 MAIN CONFIGURATION MENU ...................................................................................................3 PRINT CONFIG ...................................................................
M SERIES OPERATOR’S GUIDE LEAD GROUP DEFAULT ............................................................................................................18 PRINT 12 LEAD INTERPRETATION ............................................................................................18 DEFIB DEFAULT TO PADS .........................................................................................................18 CELL PHONE TYPE ......................................................................................
Introduction The M Series CCT is designed with several user selectable features that allow operators to configure the device according to their protocols and local requirements. This manual describes the configuration options and how to change them. This manual also lists the configurable features, options and default settings. Some features are available to manual mode operators only. The default settings shown below are for North American units only.
M SERIES CCT CONFIGURATION GUIDE Select Language 6\VWHP &RQILJXUDWLRQ 6HOHFW /DQJXDJH (1*/,6+ )5$1&$,6 3UHY 9DOXH 1H[W 9DOXH (QWHU ([LW &RQILJ The first configurable option displayed after entering the correct pass code is the Select Language option. Press the Prev Value and Next Value softkeys to move the highlight from one language to the other. Some M Series CCT units include software that allows the device to operate in either of two languages.
Configuring the Unit Main Configuration Menu The Main Configuration menu is displayed whenever the Enter softkey is depressed from the Language Select menu or the Main Menu softkey is depressed while displaying other system configuration screens. 6\VWHP &RQILJXUDWLRQ 3ULQW &RQILJ 5HVWRUH 'HIDXOWV 5HDG )URP &DUG &KDQJH &RQILJ ([LW &RQILJ This menu provides the following five (5) softkey options: Print Config, Restore Defaults, Read From Card, Change Config, and Exit Config.
M SERIES CCT CONFIGURATION GUIDE Read From Card Pressing the Read From Card softkey prompts the user to choose either a language or configuration card. The following screen will be displayed. 6\VWHP &RQILJXUDWLRQ 5HDG :KLFK 7\SH RI &DUG" /$1*8$*( &21),*85$7,21 3UHY 9DOXH 1H[W 9DOXH (QWHU To move the highlighted area from LANGUAGE to CONFIGURATION and vice versa, press the Prev Item and Next Item softkeys.
Configuring the Unit Change Config Pressing the Change Config softkey causes the unit to display the first configuration option menu. There are three user configurable features contained within each option menu. (Refer to the “Configuration Option Tables” on page 29 for all available options and default settings).
M SERIES CCT CONFIGURATION GUIDE (This page has intentionally been left blank.
Configuration Settings Configuration Settings 6\VWHP &RQILJXUDWLRQ 127&+ ),/7(5 +] $//2: &$5' (5$6(
M SERIES CCT CONFIGURATION GUIDE Set Report Restart Delays Allows review of, or changes to the length of time the device must be turned off before it erases all data stored in either summary report memory or trend memory. With Set Report Restart Delay highlighted and Set/Review selected, press the Change Value softkey.
Configuration Settings With Code Markers highlighted and Config/Review selected, press the Change Value softkey. The following screen displays: 6\VWHP &RQILJXUDWLRQ '(),% &2'( 0$5.(56 &KDQJH 5HYLHZ 021,725 &2'( 0$5.(56 &KDQJH 5HYLHZ 3$&( &2'( 0$5.(56 &KDQJH 5HYLHZ &KDQJH 9DOXH 3UHY ,WHP 1H[W ,WHP (QWHU 5HWXUQ Pressing the Change Value softkey will change the highlighted setting from Change/Review to Set to Defaults and vice versa.
M SERIES CCT CONFIGURATION GUIDE Device Identifier Sets an eleven-digit device identifier code. This code is divided into two sub codes (Site ID, eight alphanumeric characters, and Device ID, three alphanumeric characters). The device identifier is printed on the 12-lead strip chart, summary report, fax and Catalyst™ MUSE® printouts. Note: If you are transmitting 12-lead ECG data to a Catalyst MUSE system, it is recommended that you set up the device identifier.
Configuration Settings Manual Mode Passcode (Advanced AED Models) Allows the three-digit Manual Mode Passcode to be set or changed. Each digit entered can only be set between 0 (zero) and 3 (three) inclusive. Entering “000” as the pass code disables password protection of the Manual Mode. When “000” is selected, no password is required to enter the Manual Defibrillation Mode.
M SERIES CCT CONFIGURATION GUIDE Energy Level: Shock 1, 2, 3 Selects the energy levels used for the 1st, 2nd, and 3rd consecutive shocks that the defibrillator delivers when: • • In semi-automatic or advisory mode and using pads Ιn manual mode, using external paddles or pads, with Basic Auto Energy Escalation set to “Yes” Energy level 1 is the energy level the unit defaults to when the power is turned on with external paddles or pads connected.
Configuration Settings Store to Card in Auto Monitor Mode (AED models) Sets unit to automatically store patient’s ECG trace and voice data to the PCMCIA card when: • ECG leads are attached to the patient • MFE pads are not connected • M Series CCT unit is on If disabled, the M Series CCT unit will not record any data to the PCMCIA card unless MFE pads are connected to the patient.
M SERIES CCT CONFIGURATION GUIDE Pace Rate Setting on Power-Up Sets the default pace rate of the M Series CCT unit on power-up, between 30 ppm and 102 ppm (increments of 4 ppm). Display Implanted Pace Spikes The M Series CCT is capable of detecting and indicating on the display pacemaker signals from a patient with an implantable pacemaker. Selecting “Yes” causes a pacemaker marker to be displayed on the ECG trace whenever implanted pacer stimuli are detected.
Configuration Settings Fax/Communication Phone Numbers Allows the user to pre-configure up to 24 locations and phone numbers and delete obsolete phone numbers. The M Series CCT uses the phone numbers to transmit data to a fax machine or Catalyst MUSE system following 12-lead analysis. If you are transmitting to a Catalyst MUSE system, you can configure multiple phone numbers to access the same system.
M SERIES CCT CONFIGURATION GUIDE 4. Press Details softkey to select field. The Location and Phone Number menu displays: 6\VWHP &RQILJXUDWLRQ /RFDWLRQ 3KRQH 1XPEHU B B B B B B B B B B B B B B B B B B B B B B B B B B B )D[ RU 'DWD" )D[ 3UHY /HWWHU ,QF /HWWHU 'HF /HWWHU 1H[W /HWWHU (QWHU To update location of medical facility: 1. Press Prev Letter or Next Letter softkey to select letter. 2. Press Inc Letter or Dec Letter softkey to change value of letter.
Configuration Settings Deleting Fax/Communication Information The following procedure deletes obsolete phone numbers from your M Series CCT unit. To delete phone numbers: 1. From the main System Configuration menu, press Next Item softkey until “FAX/ COMMUNICATION PHONE NUMBERS” highlights. 2. Press Change Value softkey. The Phone Numbers menu displays. 3. Press Next or Prev softkey to scroll to phone number. 4. Press Delete softkey to delete phone number.
M SERIES CCT CONFIGURATION GUIDE Set Lead Groups This setting allows the user to configure two custom lead groups as described in the 12-Lead ECG Monitoring insert (part number 9650-0215-01). Pressing the Change Value softkey will cause the following screen to appear: 6\VWHP &RQILJXUDWLRQ &XVWRP *URXS ,, ,,, D9) &XVWRP *URXS ,, ,,, 9 3UHY /HDG 1H[W /HDG &KDQJH /HDG (QWHU Pressing the Prev Lead softkey will select the previous lead.
Configuration Settings Cell Phone Type This setting controls the active cellular phone profile for PCMCIA modem cards. It has no effect on landline functionality. Pressing the Change Value softkey selects different phone makes. Select the phone make corresponding to your target cellular phone. For additional information on cellular phone compatibility contact The Supply Net, Inc. at www.thesupplynet.com or (800) 826-0279.
M SERIES CCT CONFIGURATION GUIDE On M Series CCT units, you can set the NIBP cuff inflation pressure for three patient type modes: adult, pediatric and neonatal. With NIBP First Inflate Pressure Default highlighted and Set/Review selected, press the Change Value softkey.
Configuration Settings MUSE Site and Location Sets the Catalyst MUSE site and location identifiers needed to transmit 12-lead ECG data to a Catalyst MUSE system. The medical facility’s Catalyst MUSE system administrator provides these numbers when you initially set up access to the system. The Catalyst MUSE site has a value between 01 and 32, and the location has a value between 001 and 600. Note: You must also configure the Catalyst MUSE communications phone number.
M SERIES CCT CONFIGURATION GUIDE every 5, 10, 15, 30, or 60 minutes. All data samples are printed on the strip chart regardless of this setting. Trending Enabled When set to “Yes”, the M Series CCT unit will print and display trend reports as long as the appropriate hardware is installed. Defaults to “No” for units not equipped with the appropriate hardware.
Configuration Settings Alarm Limits at Power Up Allows review of, or changes to, the power-up default alarm state and alarm limits for each physiological parameter installed in the unit. With Alarm Limits At Power Up highlighted and Set/Review selected, press the Change Value softkey.
M SERIES CCT CONFIGURATION GUIDE Highlight the Clock Synchronization option and press the Change Value softkey to display the following screen: 6\VWHP &RQILJXUDWLRQ &ORFN 6\QFKURQL]DWLRQ &KDQJH 9DOXH 7LPH =RQH (67 (DVWHUQ 'LDO /RFDWLRQ &RORUDGR 'LDO 3UHIL[ B B B B B B B B 3UHY ,WHP 1H[W ,WHP 5HWXUQ Press the Change Value softkey to scroll through the available values for each parameter. Press the Next Item softkey to move to the next parameter in the table.
Configuration Settings the “Allow Card Erase” option must be set to “Yes”, and the “Record to Card in Advisory Mode” option must be set to “ECG only” or “Voice and ECG”; otherwise, this option is ignored. If set to “Yes”, the unit will display the message “Erase Card -- Yes/No?” when the user turns the Selector Switch to the OFF position and the following criteria are met: 1. A data card is installed in the M Series CTT unit. 2.
M SERIES CCT CONFIGURATION GUIDE Temperature Units Sets the temperature measurement unit to °F or °C. Display Colors Allows the user to set individual display colors for parameters displayed on the M Series CCT monitor.
Configuration Settings IBP1 Label at Power Up Sets the label assigned to the first IBP channel when the unit is powered up. IBP2 Label at Power Up Sets the label assigned to the second IBP channel when the unit is powered up. IBP1 Range at Power Up Sets the default display scale (in mmHg) for the waveform amplitude of the first IBP channel at power-up. IBP2 Range at Power Up Sets the default display scale (in mmHg) for the waveform amplitude of the second IBP channel at power-up.
M SERIES CCT CONFIGURATION GUIDE (This page has intentionally been left blank.
Configuration Option Tables Configuration Option Tables The following tables list the M Series CCT options and their possible values. An asterisk (“*”) refers to features that are currently not implemented on the M Series CCT. These features will only operate in accordance with their pre-configured default settings (listed below).
M SERIES CCT CONFIGURATION GUIDE Feature Options Default Clock Synchronization Set/Review Set/Review Time Zone Azores, Middle Atlantic, S.
Configuration Option Tables Advisory and Semi-automatic Modes Feature Options Default Energy Level: Shock 1 Damped Sine Wave (DSW): 1-10, 15, 20, 30, 50, 75, 100, 150, 200, 300, 360 J Manual DSW: 200 J Manual Biphasic: 120 J Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, 200 J AED DSW: 200 J AED Biphasic: 120 J Damped Sine Wave (DSW): 1-10, 15, 20, 30, 50, 75, 100, 150, 200, 300, 360 J Manual DSW: 300 J Manual Biphasic: 150 J Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, 200 J AED DSW:
M SERIES CCT CONFIGURATION GUIDE Manual Mode Feature Options Default Auto Generate Strips Yes/No Yes Retain SYNC after Defib Yes/No No Print 3 Leads When Leads Sel.
Configuration Option Tables EtCO2 Option Feature Options Default EtCO2 Options Set/Review Set/Review Default EtCO2 Units mmHg, kPa, % mmHg Displayed Zoom Level (in mmHg): 0-12.5, 0-25, 0-50, 0-75, 0-100 (in kPa): 0-1.7, 0-3.3, 0-6.6, 0-10.0, 0-13.3 0-50 mmHg 0-6.6 kPa Default EtCO2 Comp Setting None, O2, N2O, O2 & N2O None Feature Options Default NIBP Units (Blood Pressure Units on CCT models) mmHg, kPa mmHg NIBP Auto Interval Default 2.
M SERIES CCT CONFIGURATION GUIDE IBP Option Feature Options Default Blood Pressure Units mmHg, kPa mmHg IBP1 Label at Power Up P1, ART, PA, CVP, ICP P1 IBP2 Label at Power Up P2, ART, PA, CVP, ICP P2 IBP1 Range at Power Up (in mmHg): 0-300, 0-180, 0-120, 0-60, 0-30 (in kPa): 0-40, 0-24, 0-16, 0-8, 0-4 0-300 mmHg IBP2 Range at Power Up (in mmHg): 0-300, 0-180, 0-120, 0-60, 0-30 (in kPa): 0-40, 0-24, 0-16, 0-8, 0-4 0-120 mmHg Feature Options Default Temperature Units °C, °F °F Tempera