CPR Parameters
9650-1201-01 Rev. M 61
CPR Parameters
The CPR parameters are attributes that relate to the E Series Real CPR Help
®
functionality and
include:
• Display CPR Compression Bar Graph
• CPR Record Data to Card
• CPR Guidance Audio Prompts
• CPR Compression Metronome
• Target CPR Depth
• Display Fully Release Prompt
Display CPR Compression Bar Graph
When set to Yes, this parameter specifies that the CPR compression bar graph will
automatically replace the second waveform on the E Series display whenever chest
compressions are detected.
Values are as follows:
CPR Record Data to Card
This parameter, when set to Yes, causes CPR data to be recorded to the PCMCIA card
whenever CPR-D•padz
®
are in use. This parameter can be set to Yes only if Record ECG Data
in Advisory Mode is set to Yes.
Values are as follows:
CPR Guidance Audio Prompts
This parameter allows you to select the modes in which CPR-related audio prompts will be
issued, Semi-Auto Only (default), Yes (Manual and Semi-Auto), or No (no prompts regardless
of mode). In Semi-Auto Only, or Yes, the audio prompts are:
• PUSH HARDER when the detected compression depth is consistently less than the target
CPR depth.
• GOOD COMPRESSIONS when the rescuer responds by increasing compression depth to
the target CPR depth or more.
Values are as follows:
Possible Values: Yes, No
Default: Yes
Possible Values: Yes, No
Default: No
Possible Values: Yes, Semi-Auto Only, No
Default: Semi-Auto Only