CPR Parameters
9650-1201-01 Rev. M 61
CPR Parameters
The CPR parameters are attributes that relate to the E Series Real CPR Help
®
functionality and
include:
Display CPR Compression Bar Graph
CPR Record Data to Card
CPR Guidance Audio Prompts
CPR Compression Metronome
Target CPR Depth
Display Fully Release Prompt
Display CPR Compression Bar Graph
When set to Yes, this parameter specifies that the CPR compression bar graph will
automatically replace the second waveform on the E Series display whenever chest
compressions are detected.
Values are as follows:
CPR Record Data to Card
This parameter, when set to Yes, causes CPR data to be recorded to the PCMCIA card
whenever CPR-Dpadz
®
are in use. This parameter can be set to Yes only if Record ECG Data
in Advisory Mode is set to Yes.
Values are as follows:
CPR Guidance Audio Prompts
This parameter allows you to select the modes in which CPR-related audio prompts will be
issued, Semi-Auto Only (default), Yes (Manual and Semi-Auto), or No (no prompts regardless
of mode). In Semi-Auto Only, or Yes, the audio prompts are:
PUSH HARDER when the detected compression depth is consistently less than the target
CPR depth.
GOOD COMPRESSIONS when the rescuer responds by increasing compression depth to
the target CPR depth or more.
Values are as follows:
Possible Values: Yes, No
Default: Yes
Possible Values: Yes, No
Default: No
Possible Values: Yes, Semi-Auto Only, No
Default: Semi-Auto Only