Chapter 2 Configuration Parameters
48 9650-1201-01 Rev. M
Custom 5 Wire Grouping in Use
This parameter allows you to specify whether you are using a standard (default) ECG lead
group or a custom 5 wire grouping (II, III, aVF). When this parameter value is set to No, leads
are printed as follows:
When this parameter is set to Yes, leads are printed as follows:
Values are as follows:
Selected Lead Leads Printed (top to bottom)
I I, II, III
II II, I, III
III III, I, II
aVR aVR, aVL, aVF
aVL aVL, aVR, aVF
aVF aVF, aVR, aVL
V1 (5 Wire Cable) V1, I, aVR
Selected Lead Leads Printed (top to bottom)
II II, III, aVF
III III, II, aVF
aVF aVF, II, III
Possible Values: Yes, No
Default: No