Chapter 2 Configuration Parameters
28 9650-1201-01 Rev. M
Pace Parameters
The Pace parameters are attributes that control functions related to performing pacing with an
E Series unit and include:
Pace Rate Setting on Power Up
Async Softkey in Pace Mode
Pace Rate Setting on Power Up
This parameter allows you to set the default pace rate, pulses per minute (ppm), of the E Series
unit at power-on.
Values are as follows:
Async Softkey in Pace Mode
This parameter allows you to set the unit to display the Async Pacing On/Off softkey when the
unit is in pace mode. For information about Asynchronous Pacing, refer to the E Series
Operators Guide.
Values are as follows:
Possible Values: 30 ppm to 102 ppm (inclusive in increments of 4 ppm)
Default: 70 ppm
Possible Values: Yes, No
Default: Yes