Advisory Defibrillation Parameters
The Advisory Defibrillation parameters control functions related to performing defibrillation
with an E Series unit in semiautomatic and advisory mode and include:
• Energy Level: Shock 1
• Energy Level: Shock 2
• Energy Level: Shock 3
• Auto Charge in Advisory Mode
• Auto Analyze
• Stacked Shocks
• Display DO CPR Message
• CPR Message After Shock
• CPR Message After No Shock Advised
• Duration of CPR Interval
• Check Pulse
• Restart Analysis After CPR
• Check Patient Prompt
• Enable Voice in Advisory Mode
• Continual Check Patient Audio Prompt
• Enable SCE
• Post SCE Result on Display
Energy Level: Shock 1
This parameter allows you to select the energy level used for the first shock delivered by the
defibrillator when the unit is:
• in semi-automatic or advisory mode and using pads
• in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes
Energy Level: Shock 1 is the energy level to which the unit defaults when power is turned on
with external paddles or pads connected.
Values are as follows:
Energy Level: Shock 2
This parameter allows you to select the energy level used for the second shock delivered by the
defibrillator when the unit is:
• in semi-automatic or advisory mode and using pads
• in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes
Possible Values: 1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J
Default: 120 J