Owner manual

E Series - Pulse CO-Oximetry (SpO
2
, SpCO, and SpMet) Option Insert
10 9650-1211-01 Rev. H
Default Settings
When the pulse CO-oximeter is turned on, the following default settings are automatically selected and remain in
operation until changed.
Note: Only the default settings are available in AED Semi-Automatic operation. Default alarm limit setting may be
adjusted in System Configuration mode. See the E Series Configuration Guide for more information.
CO-Oximeter Accessories
The following table describes each of the SpO
2
, SpCO, and SpMet accessories.
Note: LNCS single use and reusable sensors are capable of measuring SpO
2
only. Rainbow single use sensors are
capable of measuring SpO
2
, SpCO, and SpMet.
Parameter Default Setting Range
Averaging Mode 8 seconds (medium) 4 seconds (short)
8 seconds (medium)
16 seconds (long)
Sensitivity Normal Normal or High
High SpO
2
Alarm Limit 100% 72% to 100%
Low SpO
2
Alarm Limit 85% 70% to 98%
High SpCO Alarm Limit 10% 2% to 100%
Low SpCO Alarm Limit 0% 0% to 99%
High SpMet Alarm Limit 3.0% 1% to 100%
Low SpMet Alarm Limit 0.0% 0% to 99%
High Heart Rate Alarm Limit 150 bpm (beats/minute) 60 to 280 bpm (beats/minute) - monitoring via
ECG
60 to 235 bpm (beats/minute) - monitoring via
Pulse Oximetry
Low Heart Rate Alarm Limit 30 bpm (beats/minute) 20 to 100 bpm (beats/minute)
Item Description REF
LNCS Adtx Single use sensor for patients > 30 kg 8000-0320
LNCS Pdtx Single use sensor for Pediatrics and Slender Adults 10 - 50 kg 8000-0321
LNCS Inf-3 Single use sensor for Infants 3 - 20 kg 8000-0322
LNCS Neo-3 Single use sensor for Neonates < 3 kg 8000-0323
LNCS NeoPt-3 Single use sensor for Neonates < 1 kg (Pre-term) 8000-0324
LNCS DCI Reusable sensor for Adults and Pediatrics > 30 kg 8000-0294
LNCS DCIP Reusable sensor for Pediatrics 10 - 50 kg 8000-0295
Red LNC-04 4' Reusable Patient Cable - connects to LNCS single use and reusable
sensors listed above for monitoring SpO
2
.
8000-0330
Red LNC-10 10' Reusable Patient Cable - connects to LNCS single use and reusable
sensors listed above for monitoring SpO2.
8000-0331