User Manual
E Series - End Tidal Carbon Dioxide (EtCO
2
) Option Insert
20 9650-1212-01 Rev. E
Software Hazards Minimized by compliance with EN1441
Operating Time
EtCO
2
and SpO
2
Options
For a new fully charged PD4410 battery
pack at 20º C (68° F):
• 35 defibrillator discharges at maximum
energy (200 J), or
• 1.5 hours minimum of continuous ECG
monitoring, or
• 1.0 hour of continuous ECG monitoring/
pacing at 60 mA, 70 beats per minute.
For a new, fully charged lithium ion battery
pack at 20°C (68°F):
• 90 defibrillator discharges at maximum
energy (200 J), or
• 3.0 hours minimum of continuous ECG
monitoring, or
• 2.5 hours of continuous ECG
monitoring/pacing at 60 mA, 70 beats
per minute.
CAPNOSTAT 5® CO
2
Sensor LoFlo™