User Manual

E Series - End Tidal Carbon Dioxide (EtCO
2
) Option Insert
14 9650-1212-01 Rev. E
Default Settings
When the unit is turned on, the following default EtCO
2
settings are automatically selected and remain in operation
until changed.
Note The power-on default settings for the capnogram waveform scale and CO
2
compensation are set in System
Configuration, as are the power-on default settings for alarm limits. See the E Series Configuration Guide for
more information.
EtCO
2
Accessories
The following table lists the accessories available for the E Series mainstream EtCO
2
monitoring option.
Table 1: CAPNOSTAT 5 Mainstream CO
2
Accessories
*
SPU = Single Patient Use
Parameter Default Setting Range
Averaging Mode 10 seconds 1 breath
10 seconds
20 seconds
High EtCO
2
Alarm Limit 55 mmHg
7.2%
7.3 kPa
5 - 100 mmHg, OFF
0.6 - 13.1%, OFF
0.6 -13.3 kPa, OFF
Low EtCO
2
Alarm Limit 25 mmHg
3.2%
3.3 kPa
0 - 95 mmHg, OFF
0 - 12.5%, OFF
0-12.6kPa, OFF
High Respiration Rate Alarm Limit 120 respirations per minute 5 - 150 respirations per minute, OFF
Low Respiration Rate Alarm Limit 5 respirations per minute 0 - 100 respirations per minute, OFF
Accessory REF Application
CAPNOSTAT 5 CO
2
Sensor and Cable 8000-0312-01
SPU
*
Pediatric/Adult Airway Adapter 8000-0260-01 Single patient use, for ET tube sizes > 4.0 mm
SPU
*
Neonatal/Pediatric Airway Adapter 8000-0261-01 Single patient use, for ET tube sizes < 4.0 mm
Reusable Adult Airway Adapter 8000-0262-01 Reusable, for ET tube sizes > 4.0 mm
Reusable Neonatal/Pediatric Airway Adapter 8000-0263-01 Reusable, for ET tube sizes <
4.0 mm
SPU
*
Pediatric/Adult Airway Adapter with
Mouthpiece
8000-0265-01 Single patient use, for non-intubated patients
CAPNO
2
mask, Large Adult 8000-0761 SPU, for non-intubated large adults
CAPNO
2
mask, Standard Adult 8000-0760 SPU, for non-intubated adults
CAPNO
2
mask, Pediatric 8000-0762 SPU, for non-intubated adults pediatric patients