E Series Service Manual
9650-1450-01 Rev. D 2-9
ECG LEAD OFF One or more ECG leads are not properly connected when
leads are selected as input.
Check cable and patient connection. Change
electrodes. Prepare patient’s skin.
ECG TOO LARGE ECG signal too large for accurate shockable rhythm
analysis.
Reduce ECG size.
ECG V LEAD OFF V LEAD cable is not properly attached.
Attach V LEAD cable.
ECG V1 LEAD OFF Chest lead V1 is not properly attached to patient.
Reattach V lead. Check cable.
ECG V2 LEAD OFF Chest lead V2 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V3 LEAD OFF Chest lead V3 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V4 LEAD OFF Chest lead V4 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V5 LEAD OFF Chest lead V5 is not properly attached to patient.
Reattach V lead.
Check cable.
ECG V6 LEAD OFF Chest lead V6 is not properly attached to patient.
Reattach V lead. Check cable.
ENTER ACCESS CODE Manual mode access code needed.
Enter access code to enter manual mode with
AED.
ERASING REPORT Summary report being erased.
FAX BUSY FAX line at receiving FAX is busy.
FAX DIALING Preparation for sending fax.
FAX DONE Transmission complete.
FAX PREPARING Preparing fax for transmission.
FAX SENDING Transmitting fax.
Error Message Explanation User
Advisory
Technical Action