E Series Service Manual
2-8
9650-1450-01 Rev. D
DEFIB FAULT 109 Defib capacitor voltage is greater than selected energy
when defibrillator is charging or ready.
Replace high voltage module or capacitor.
DEFIB FAULT 111 Defib capacitor voltage has exceeded the absolute
maximum acceptable voltage.
Replace high voltage module, capacitor, and or
system board.
DEFIB FAULT 195 Current higher than expected was detected during the
Biphasic bridge test or immediately following a discharge.
Ensure pads/paddles are used properly.
Attempt to clear the message by turning the
Selector switch to off then back to the desired
operating modes.
Replace bridge or high voltage module.
DEFIB FAULT 196 Biphasic module not operating properly while charging.
Charge again.
Attempt to clear the message by turning the
Selector switch to OFF, then back to the desired
operating mode.
Replace bridge or high voltage module.
DEFIB NOT CHARGED Discharge button is pressed but the unit is not charged.
DEFIB PAD SHORT Measured impedance between high voltage leads of MFC.
Ensure pads are coupled to patient. Check /
replace pads or universal cable.
Replace system board.
DISABLE SYNC Sync mode active when analyze pressed in defib.
DISCHARGE FAULT Defib capacitor voltage is not decreasing. Replace high voltage module, capacitor, and/or
system board.
ECG FAULT 4 Communication fault between ECG processor and main
processor.
Turn off unit and then turn on to reset. If fault
persists, replace system board.
ECG FAULT 5 ECU RAM test failure, or ROM checksum test failure. Turn off unit and then turn on to reset. If fault
persists, replace system board.
Error Message Explanation User
Advisory
Technical Action