Owner manual
Do this... Verify that... Pass/Fail
23.1
23.2
o o
23.3
• Systolic pressure (115 - 125 mmHg)
• Diastolic pressure (75 - 85 mmHg)
• Mean pressure (88 - 98 mmHg)
o o
23.4
o o
E Series Service Manual
1-40 9650-1450-01 Rev. D
4
These values only apply for test set-ups utilizing the BP Pump 2 Simulator. Variations of the test set-up or different simulators may produce readings
outside
the provided values and will require end-user facility to establish the appropriate offset and tolerances.
Please
contact ZOLL Technical Support i f you
require
assistance establishing the offset o f y o u r simulator and test set-up
.
Turn the selector switch to MONITOR mode.
(For AED units, turn the selector switch to ON
and select Manual mode.)
Ensure that the LEADS p
arameter is set to
PADS (default).
If necessary, press the LEADS button to cycle
through the values to select PADS.
The E Series unit displays PADS in the Lead selection field on the monitor.
Press the NIBP bu
tton on the E Series front
panel.
The E Series displays the following measurements:
Press the P
ARAM softkey, then press the
TREND softkey, followed by the NIBP Trend
softkey.
The E Series unit displays a summary of the NIBP measurements, including
the pulse rate reading (in the range of 77 - 83 bpm).
Record your results on the Maintenance Tests Checklist.