1-16 9650-1450-01 Rev. D
8.0 Calibrating Pulses on Strip Chart Test
Tools Needed: None
Test Setup: No
ne.
Do this... Verify that... Pass/Fail/NA
8.1
8.2
o o o
Press the RECORDER button.
Press and hold SIZE but
ton to activate the
calibration signal.
The strip chart displays a signal of 300 ppm with an amplitude of
10 mm +/- 1 mm. The signal also appears on the video display.
Record your results on the Maintenance Tests Checklist.