7.0 Heart Rate Display Test
Tools Needed:
• ECG Cable (3 or 5 leads).
• QED 6 Defibrillator Analyzer.
Test Setup:
1. Turn the selector switch to MONITOR. Press LE
AD button until ā€œIā€ displays.
2. Connect the ECG leads to the QED 6 Defibrillator Analyzer.
3. Connect the ECG cable to the unit.
Do this... Verify that... Pass/Fail/NA
7.1
o o o
Set the QED 6 Defibrillator Analyzer to
120BPM.
The Heart Rate displays as 120 +/- 2 bpm
Record your results on the Mai