Owner manual
Troubleshooting Guidelines
9650-1210-01 Rev. R 13-7
AC Charger
Symptom Recommended Action
The green and orange-yellow CHARGER ON indicators
illuminate alternately.
• Verify battery is installed.
• Turn unit ON to identify the fault condition.
• Replace battery pack with a fully charged battery pack.
• If problem persists, replace battery pack, unplug device
from AC mains and plug device back into AC mains.
Unit displays the LOW BATTERY message when it is
plugged into AC mains.
• Replace battery pack with a fully charged battery pack.
• Unplug device from AC mains and plug device back into
AC mains.
• Verify AC mains is working properly.
None of the CHARGER ON indicators are illuminated
when the device is plugged into AC mains.
• Unplug device from AC mains and plug device back into
AC mains.
• Verify AC mains is working properly.