9650-1450-01 Rev. D 1-13
5.0 Leakage Current Test
Tools Needed: See the manufacturers instructions or supplied specifications for the leakage tester you use.
Test Setup: Se
e the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with accessories: MFC,
external paddles, and anterior/posterior paddles.
Maximum Leakage Acceptance Limits
*Single fault considered AC mains on applied part.
Normal Condition Single Fault Condition*
ECG 10 50
MFC 100 100
Earth 500 1000