User Manual

P/N. 10762-001 Rev. 8 Page B-3
Battery Charger User Guide
B.5 FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
B.6 Limited Warranty for AutoPulse Resuscitation System
ZOLL Circulation warrants to the initial Purchaser only that the “Warranted Product” purchased
hereunder will be free from defects in workmanship or materials, when given normal, proper, and
intended usage, for a specied period (“Warranty Period”) from the date of its initial shipment to
Purchaser. “Warranted Products” consist solely of those products whose description in this price list
Shock Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)
Vibration Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s
2
)
Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
Free fall Meets IEC 60068-2-32 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.
Safety requirements Safety certied to UL2601, CSA 22.2 No. 601.1-M90, EN60601-1
Table B-6 Battery Charger EMI/EMC Specifications
Title Standard
Electrostatic discharge IEC 61000-4-2, Level 3
RF electromagnetic elds
immunity
IEC 61000-4-3, Level 2
EFT/burst IEC 61000-4-4, Level 2
Surge immunity IEC 61000-4-5, Level 2
Conducted RF disturbances
immunity
IEC 61000-4-6, Class A
Dips, interruptions, and
variations
IEC 61000-4-11
Harmonics current emissions IEC 61000-3-2, Class A
Radiated emissions CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A
Table B-5 Battery Charger Environmental Specifications (Page 2 of 2)
Category Specifications