Owner manual

P/N. 12457-001 Rev. 2 Page C-3
Power System User Guide
C.3 Multi-Chemistry Battery Charger Physical And
Environmental
Free fall Meets IEC 60068-2-31 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.
Electrostatic discharge Meets IEC 61000-4-2, Level 3
Radiated emissions Meets CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A
Table C-3 Multi-Chemistry Battery Charger Specifications (Page 1 of 2)
Category Specifications
Manufacturer Manufactured for ZOLL Circulation
Model Number 8700-0753-01
Size (LWH) 16.01 in. by 9.50 in. by 6.54 in. (40.6 cm by 24.1 cm by 16.6 cm).
Weight 7.1 lbs. (3.23 kg)
Operating input voltage 100 to 240V AC
Operating input frequency 50/60 Hz
Input current 2.0 Amps (maximum)
Maximum Battery charge
time
Less than 6¼ hours (at 77ºF [25ºC])
Fuses User-replaceable, T2.5A 250V AC (2 required)
Operating temperature +41º to +95ºF (5º to +35ºC) (68º to 77ºF [20º to 25ºC] preferred)
Storage temperature -40º to +158ºF (-40º to +70ºC)
Relative humidity 5% to 95%, non-condensing.
Operating altitude 0 to 10,000 ft. (0 to 3,048 m)
Enclosure protection Meets IP22 per IEC 60529
Shock Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)
Vibration Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s
2
)
Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
Free fall Meets IEC 60068-2-31 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.
Electrostatic discharge Meets IEC 61000-4-2, Level 3
Table C-2 NiMH Battery Specifications (Page 2 of 2)
Category Specifications