Owner manual
P/N. 12457-001 Rev. 2 Page C-1
Power System User Guide
Appendix C Technical Specifications
The specifications provided in this appendix apply to the AutoPulse Power System.
C.1 Li-Ion Battery Physical and Environmental
Table C-1 Li-Ion Battery Specifications (Page 1 of 2)
Category Specifications
Manufacturer ZOLL Circulation
Model Number 8700-0752-01
Size (LWH) 11.5 in. by 3.2 in. by 2.2 in. (29.2 cm by 8.1 cm by 5.7 cm).
Weight 3.0 lbs. (1.3 kg).
Type Rechargeable Lithium-Ion (LiFePO
4
)
Battery voltage (nominal) 36.3V
Capacity 2300 mAh (typical)
Current (maximum) 30 A continuous, 48 A pulse (96 ms max)
Initial Battery run time
(nominal patient)
30 minutes (typical)
Maximum Battery charge
time
Less than 4¼ hours at 77ºF (25ºC)
Battery Test-Cycle time Less than 12 hours per Test-Cycle session
Recommended replacement
interval
3 years from date of manufacture
Note: The Battery will not operate after 5 years from date of manufacture.
Operating temperature +32º to +113ºF (0º to +45ºC) ambient installed in device
Charge temperature +41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred)
Storage temperature -4º to +113ºF (-20º to +45ºC) ambient for up to six months with charging every
four weeks, starting with a fully charged Battery.
Operating altitude 0 to 15,000 ft. (0 to 4,572 m)
Enclosure protection Meets IP24 per IEC 60529
Shock Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)
Vibration Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s
2
)
Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
Free fall Meets IEC 60068-2-31 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.
Electrostatic discharge Meets IEC 61000-4-2, Level 3