Owner manual

Page 7-2 P/N. 12457-001 Rev. 2
Power System User Guide
3. Check both of the fuses. If a fuse must be replaced, follow these steps:
a) Replace both of the original fuses with T2.5A 250V AC fuses.
b) Push in the fuse holder until the locking tab clicks into place.
c) Re-attach the Power Cord.
d) When the AutoPulse Multi-Chemistry Battery Charger is powered on, all the LED’s illuminate
briefly as the Charger performs a self test.
Note: While the self test is being performed, if the indicator lights remain illuminated or if one
of the indicator lights fails to illuminate, contact ZOLL.