36 Fully Automatic AED Plus Administrator’s Guide
Figures 9 through 14 show the Rectilinear Biphasic waveforms that the AED Plus defibrillator produces
when it discharges into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy setting (200,
150, 120, 85, 70, and 50 joules).
The vertical axis shows the current in amperes (A); the horizontal axis shows the duration in
milliseconds (ms).
Figure 9: Rectilinear Biphasic Waveforms at 200 joules
Figure 10: Rectilinear Biphasic Waveforms at 150 joules
-30
-25
-20
-15
-10
-5
0
5
10
15
20
25
30
35
0 2 4 6 8 10 12 14 16 18 20
Current (A)
Time (ms)
Patient
Impedance
25ohms
50ohms
75ohms
100oh ms
125oh ms
150oh ms
175oh ms