Owner manual

Fully Automatic AED Plus Administrator’s Guide 29
Appendix A: Specifications
Table 7: General Specifications
DEVICE
Size (H x W x D) 5.25" x 9.50" x 11.50"; 13.3 cm x 24. 1 cm x 29.2 cm
Weight 6.7 lbs.; 3.1 kg
Power User Replaceable Batteries. 10 Type 123A Photo Flash lithium
manganese dioxide batteries
Device Classification Class II and internally powered per EN60601-1
Design Standards Meets applicable requirements of UL 2601, AAMI DF-39,
IEC 601-2-4, EN 60601-1, IEC 60601-1-2
ENVIRONMENT
Operating Temperature PS Model: 32° to 122° F; 0° to 50° C
Storage Temperature PS Model: -22° to 158° F; -30° to 60° C
Humidity 10 to 95% relative humidity, non-condensing
Vibration MIL Std. 810F, Min Helicopter Test
Shock PS Model: IEC 68-2-27; 100G
Altitude PS Model: -300 to 15,000 ft.; -91m to 4573m
Aircraft Method RTCA/DO-160D: 1997
Section 21, Category M – all operating modes.
Particle and Water Ingress IP-55
DEFIBRILLATOR
Waveform Rectilinear Biphasic™
Defibrillator Charge Hold
Time
30 seconds
Energy Selection
Automatic pre-programmed selection
(Adult mode: 120J, 150J, 200J; Pediatric mode: 50J, 70J, 85J)
Patient Safety
All patient connections are electrically isolated.
Charge Time
Less than 10 seconds with new batteries.
Maximum time from first rhythm
analysis to unit charged and ready
to shock
With new batteries: 12 seconds
With batteries depleted by 15 200J discharges: 13 seconds