5. ProductWorkingPrinciple
Figure2,Productworkingprinciplediagram
Theprincipleofdripspeedmeasurement:theinfraredemissiontubeiscontrolledby
theMCU,andtheinfrarediscontinuouslytransmittedtosendinfrared.Normally,the
infraredreceptiontubecanreceivethesignaloftheinfraredemissiontube.Whenthe
medicinedrops,the waterdropswill block the infrared reception. At this time, it is
judgedthatthereisadrop,andtheMCUwillcalculatethedropspeedaccordingto
theblockingtimebetweenthetwodrops.
Theprinciple ofinfusion completion:the principle ofcapacitiveinductionis usedto
completetheinfusion,andtheinductioncapacitorisplacedontheupperendofthe
Capacitivesensor,sensing
whether there is liquid in
theinfusiontube.
Cut‐off motor, stuck
when the infusion is
completed
MedicineDrop
Infraredemissiontube
InfraredReceptiontube
Medicine
LCD
Display
PowerButton
Chargeable
Battery