User's Manual
13
Guidanceandmanufacturer'sdeclaration‐electromagneticimmunity
Thisproduct(E‐SY01)isexpectedtobeusedintheelectromagneticenvironmentspecifiedbelow.
Thepurchaserorusershouldensurethatitisusedinthiselectromagneticenvironment:
ImmunityTest
IEC60601
TestLeve l
Conforming
Level
Electromagneticenvironment‐guide
ConductedRF
GB/T17626.6
RadiatedRF
GB/T17626.3
3Vrms
150kHzto
80MHz
3V/m
80MHzto
2.5GHz
3V
3V/m
Portable and mobile RF communication
equipmentshouldnotbeusedcloserto
anypartofthisproduct(E‐SY01)than
the recommended isolation distance,
includingcables.Thedistanceshouldbe
calculatedbytheformulacorresponding
tothetransmitterfrequency.
Recommendedisolationdistance
d=1.2√P
d=1.2√P80MHz~800MHz
d=2.3√P800MHz~2.5GHz
Amongthem,Pisthemaximumoutput
ratedpowerofthetransmitterprovided
by the transmitter manufacturer, in
watts (W), d is the recommended
isolationdistance,inmeters(m).
The field strength of the fixed RF
transmitter is determined by surveying
the electromagneticfield, andit should
be lower than the compliance level in
eachfrequencyrange.
Interferencemayoccurinthevicinityof
equipment marked with the following
symbol.
Note1:At80MHzand800MHz,thehigherfrequencybandformulaisused.
Note2:Theseguidelinesmaynotbesuitableforallsituations.Electromagneticpropagation
isaffectedbyabsorptionandreflectionfrombuildings,objectsandpeople.