User's Manual
11
Guidanceandmanufacturer'sdeclaration‐Electromagneticimmunity
Thisproduct(E‐SY01)isexpectedtobeusedintheelectromagneticenvironmentspecifiedbelow,
thepurchaserorusershouldensurethatitisusedinthiselectromagneticenvironment:
ImmunityTest IEC60601TestLevel Conforminglevel
Electromagnetic
environmentguide
Electrostatic
discharge
GB/T17626.2
±6kV
Contactdischarge
±8kV
Airdischarge
±6kV
Contactdischarge
±8kV
Airdischarge
The floor should be
wood,concreteor
ceramicaltiles.Ifthe
flooriscoveredwith
materials,therelative
temperatureshouldbe
atleast30%.
Electricalfast
transientburst
GB/T17626.4
±2kV
topowerline.
±1kV
oninput/outputline.
±2kV
topowerline.
Notapplicable
Thenetworkpower
supplyshouldhave
qualityusedina
typical commercial or
hospitalenvironment.
Surge
GB/T17626.5
±1kV
linetoline.
±2kV
linetoground.
±1kV
linetoline.
±2kV
linetoground.
Thenetworkpower
supplyshouldhavethe
qualityusedina
typical commercial or
hospitalenvironment.
Voltagedips,
short
interrup‐tions
and vol‐tage
changesonthe
power
inputline.
GB/T17626.11
<5%UTfor0.5cycle
(OnUT,>95%drop)
40%UTfor5cycles
(OnUT,60%drop)
70%UTfor25cycles
(OnUT,a30%drop)
<5%UTfor5s
(OnUT,>95%drop)
<5%UTfor0.5cycle
(OnUT,>95%drop)
40%UTfor5cycles
(OnUT,60%drop)
70%UTfor25cycles
(OnUT,a30%drop)
<5%UTfor5s
(OnUT,>95%drop)
Thenetworkpower
supplyshouldhave
thequalityusedin
atypicalcommercial
hospitalenvironmentIf
theuserofthis
product(E‐SY01)needs
toruncontinuously
during the power
interruption,itis
recommendedthat
this product (E‐SY01)
useuninterruptible
Powersupplyor
batterypowerSupply.