Technical data

94 Description
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
6 Flap
The flap is the mechanical indicator for the operating status of the hal-
ogen lamps.
When the flap is closed, the main lamp is operative (green light
(15) is on).
When the flap is open, the main lamp has failed. The backup lamp
is used (yellow light (14) is on).
7 Filter selector knob for the additional integrated halogen light source
The filter knob has four positions:
8 Opening the lamp module of the additional integrated halogen light
source
When you press this button, the lamp module is slightly ejected. Pull
out the lamp module all the way for lamp change.
9 Manual activation of the halogen backup lamp
If the automatic activation system fails, press this button to switch on
the backup lamp.
10 Indicator: Superlux Eye backup lamp is in use
When the red segment in the knob lights up, the backup lamp is in use.
11 Ventilation grid for the additional integrated halogen light source
Do not cover the ventilation grid! For example, drapes could be cov-
ering the grid. This can lead to overheating of the lamp modules and
to lamp failure.
12 Lamp module of the additional integrated halogen light source
13 Superlux Eye lamp module
no filter
blue barrier filter (retina protection filter): when operating on
the eye, always use this filter to protect the patient's eye
against unnecessary radiation (retinal injury).
KK 40 filter:
to increase the color temperature
empty filter position