Technical data

G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Description 93
When you press this button, the lamp module is slightly ejected. For
changing the lamp, pull out the lamp module as far as it will go. Turn
knob (5) through 180° until it snaps in. This moves the backup lamp
into the illumination beam path.
5 Manual activation of the Superlux Eye backup lamp
When the xenon lamp fails, open the lamp module as follows:
Press button (4). The lamp module is slightly ejected. Pull out the lamp
module as far as it will go. Turn knob (5) through 180° until it snaps in.
This moves the backup lamp into the illumination beam path. Push the
lamp module all the way back into the lamp housing. Use a pointed ob-
ject and press it into the recess of reset button (3).
Note:
When inserting a new lamp module, make sure that knob (2) is set to
"1“. If the first lamp fails, you switch to the second lamp in logical se-
quence.
241 53