Technical data

92 Description
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Superlux Eye with integrated halogen light source (option)
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
Replace the xenon lamp in good time.
Reset the service hour counter to "0" after replacing the lamp.
Please observe the procedure described on page 80.
Superlux Eye with an integrated halogen light source is a combination of
the two light sources described above. Like the Superlux Eye light source,
this combination is also available with two filter versions: the HaMode filter
and the 485 mm fluorescence excitation filter.
Ventilation grid
Do not cover the ventilation grid! For example, drapes could be covering
the grid. This can lead to overheating of the lamp modules and to lamp
failure.
1 Filter selector knob
The filter selector knob has the following positions:
2 Counter
The counter records the service hours of the xenon light source.
Change the xenon lamps after about 500 hours of operation to avoid
any explosion of the xenon lamps, and reset the counter to "0". Use a
pointed object and press it into the recess of reset button (3).
3 Reset button
The reset button resets the service hour counter to "0".
4 Opening the Superlux Eye lamp module
no filter
blue barrier filter (retina protection filter)
HaMode filter (standard)
485 nm fluorescence excitation filter (option)