Technical data
86 Description
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
1 Lamp module
2 Manual activation of the backup lamp
• When the xenon lamp fails, open the lamp module as follows:
• Press button (7). The lamp module is slightly ejected.
• Pull out the lamp module as far as it will go.
• Turn knob (2) through 180° until it snaps in. This moves the backup
lamp into the illumination beam path.
• Push the lamp module all the way back into the lamp housing.
• Reset the service hour counter to "0". Use a pointed object and press
it into the recess of reset button (6).
Note:
When inserting a new lamp module, make sure that knob (2) is set to
"1“. If the first lamp fails, you switch to the second lamp in logical se-
quence.
3 Indicator: backup lamp is in use
When the red segment in knob (2) lights up, the backup lamp is in use.
4 Filter selector knob
The filter selector knob has the following positions:
no filter
blue barrier filter (retina protection filter)
HaMode filter (standard)
485 nm fluorescence excitation filter (option)