Technical data

86 Description
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
1 Lamp module
2 Manual activation of the backup lamp
When the xenon lamp fails, open the lamp module as follows:
Press button (7). The lamp module is slightly ejected.
Pull out the lamp module as far as it will go.
Turn knob (2) through 180° until it snaps in. This moves the backup
lamp into the illumination beam path.
Push the lamp module all the way back into the lamp housing.
Reset the service hour counter to "0". Use a pointed object and press
it into the recess of reset button (6).
Note:
When inserting a new lamp module, make sure that knob (2) is set to
"1“. If the first lamp fails, you switch to the second lamp in logical se-
quence.
3 Indicator: backup lamp is in use
When the red segment in knob (2) lights up, the backup lamp is in use.
4 Filter selector knob
The filter selector knob has the following positions:
no filter
blue barrier filter (retina protection filter)
HaMode filter (standard)
485 nm fluorescence excitation filter (option)