Technical data

80 Description
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Halogen light source (option)
The suspension system is equipped with a light source for fiber illumina-
tion. The lamp housing contains a backup lamp which is automatically
swung into the illumination beam path when the first lamp fails. If required,
the light source can be equipped with a second lamp housing so that two
separate light sources are available for fiber illumination.
1 Lamp module
2 Ventilation grid
Do not cover the ventilation grid! For example, drapes could be cov-
ering the grid. This can lead to overheating of the lamp modules and
to lamp failure.
3 Flap
The flap is the mechanical indicator for the operating status of the hal-
ogen lamps.
When the flap is closed, the main lamp is operative (green light (9)
is on).
When the flap is open, the main lamp has failed. The backup lamp
is used (yellow light (8) is on).
4 Manual activation of the backup lamp
If the automatic activation function fails, press this button to switch on
the backup lamp.
5 Opening the lamp module
When you press this button, the lamp module is slightly ejected. Pull
out the lamp module all the way for lamp change.
6 Filter selector knob
The filter selector knob has four positions:
no filter
blue barrier filter (retina protection filter): when operating on
the eye, always use this filter to protect the patient's eye
against unnecessary radiation (retinal injury).
KK 40 filter:
to increase the color temperature
empty filter position