Technical data
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Safety 41
4 Yellow indicator lamp
– Lights when the main lamp has failed. In addition, open flap (1) on
the lamp module indicates that the main lamp has failed. The back-
up lamp is on.
– Blinks when the backup lamp has failed.
5 Manual function
When the manual function has been activated, all electrical control
systems are disabled. The lamp brightness is automatically adjusted
to a fixed setting.
blue barrier filter (retina protection filter): use the blue bar-
rier filter when operating on the eye. It protects the patient's
retina against unnecessary (blue) radiation and permits the
radiation exposure time to be increased by factor 3.
KK 40 filter:
to increase the color temperature
empty filter position
4
3
2
1
5