Technical data

40 Safety
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Superlux Eye light source with
integrated halogen light source (option)
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
Replace the xenon lamp in good time.
Reset the service hour counter to "0" after replacing the lamp.
For the lamp change procedure and how to reset the service hour
counter, see "Switching to the backup lamp" on page 38.
Warning!
Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
Before opening the lamp housing, make sure that the system is moved
to a position where neither the patient nor the user is put at risk by fall-
ing items.
Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronics mod-
ules. Inform our service department.
1 Flap
The flap is the mechanical indicator for the operating status of the hal-
ogen lamps.
When the flap is closed, the main lamp is operative.
When the flap is open, the main lamp has failed. The backup lamp
is on.
2 Switching to the backup lamp
The lamp housing contains a backup lamp which is automatically
swung into the illumination beam path when the first lamp fails. If this
automatic function fails, you can switch on the backup lamp by
pressing this button.
3 Filter selector knob
The filter knob has four positions:
no filter