Technical data
38 Safety
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
2 Filter selector knob
The filter selector knob has the following positions, depending on the
light source used:
3 Indicator: backup lamp is in use
When the red segment in the knob (1) lights up, the backup bulb is in
use.
Note:
If the first lamp has failed and the backup lamp is in use, make sure to
have a backup lamp module ready at hand as a precaution.
4 Yellow indicator lamp
Lights when the bulb has failed, or if the lamp module is defective.
After activation and ignition of the backup bulb, the yellow indicator
lamp turns off again.
5 Manual function
When the manual function has been activated, all electrical control
systems are disabled. The bulb brightness is automatically adjusted to
a fixed setting.
Warning!
Software and hardware failure may increase the brightness of the xenon
light source, leading to retinal injury in the patient's eye.
If several successive beeps are emitted and the xenon lamp is lit when
the system is switched on, this indicates a malfunction of the xenon light
source.
• Attach a sign to the system stating it is out of order and contact our
service representative.
No filter
Blue barrier filter (retina protection filter)
HaMode filter (standard)
485 nm fluorescence excitation filter (option)