Technical data
32 Safety
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Safety devices of the suspension systems
1 Release bar
Allows non-sterile persons to release the magnetic brakes of the sus-
pension system.
2 Adjustment screw for limiting downward movement
Use this knob to set the minimum vertical working distance from the
surgical field. Check this setting before
each surgical procedure.
3 Locking knob
for the horizontal position of the suspension arm
Before removing or attaching a module (microscope, tube, etc.) move
the suspension arm into its horizontal position. Pull out the locking
knob and turn it clockwise or counterclockwise through 180°. At the
same time, slightly move the suspension arm up and down until the
lock snaps in. This prevents the suspension arm from uncontrollably
moving upward when insufficient weight is attached. After mounting
the module, unlock the suspension arm and perform the balancing
procedure.