Technical data
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
What to do in the event of malfunctions 221
Malfunctions in the video monitor
In combination with three
successive beeps and dis-
play of the error code (XXX).
– Error message during
the software check after
power-on of the suspen-
sion system.
– Error message for an in-
ternal system error.
Manual operation is pos-
sible.
Contact service dept.
specify error code and serial
no.
-
S88 floor stand only: stand
wobbles.
Floor not level. Stand base
not appropriately posi-
tioned.
Slightly turn stand base. Ar-
ticulated arm should be po-
sitioned at a right angle with
tilt axis.
-
Warning signal is emitted
when suspension system is
switched on.
Superlux Eye light source is
defective.
Read chapter "Malfunction
of the Superlux Eye light
source“. If the malfunction
cannot be corrected, attach
a sign to the system stating
it is out of order and contact
the service dept.
-
Failure of the monitor Line power failure. Contact in-house electri-
cian.
-
Failure of suspension
system electronics.
Use the tube for viewing the
field of view. Contact
service dept.
-
Problem Possible cause Remedy See
XXX
Problem Possible cause Remedy See
No imge display. Main switch is switched off. Switch on the main switch. -
Power cable not correctly
plugged in.
Plug in the power cable. -
Displayed message
"NO SIGNAL"
No signal source con-
nected.
Connect a signal source. -
No camera connected. Connect a camera. -
Connected video cable is
defective.
Connect an intact video
cable.
-