Technical data

G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
What to do in the event of malfunctions 219
Malfunctions in the S8, S81 or S88 suspension system
Zoom system inoperative. Zoom system does not
work, or moves to upper or
lower end position.
Manually adjust magnifica-
tion using zoom knob.
Contact service dept.
page 215
Zoom and focusing systems
inoperative.
Functions not correctly set
on suspension system.
Contact service dept. --
Microscope motion too stiff. Friction adjusting knob on
suspension system tight-
ened too firmly.
Slightly loosen friction ad-
justing knob.
page 144
DeepView inoperative DeepView is on mechanical
end stop.
Press DeepView button to
deactivate the function.
--
Problem Possible cause Remedy See
Problem Possible cause Remedy See
No function at all. Line power failure. Contact in-house electri-
cian.
-
Power switch of suspension
system not switched on.
Press power switch. -
Automatic circuit breaker in
power switch of suspension
system has been activated.
Press power switch again. -
Surgical field illumination on
microscope not working.
Thermal cut-out activated. Remove the cause of over-
heating. For example,
drapes could be covering
the grid. When the lamp
module has cooled down,
the light source switches on
again.
-
Selector set in such a way
that illumination can be
switched on using foot con-
trol panel.
Switch on illumination using
foot control panel (button A
or B).
-
Failure of suspension
system electronics.
Illuminate surgical field
using an OR illuminator.
Contact service dept.
-