Technical data

216 What to do in the event of malfunctions
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Causes of malfunctions and remedies
For your safety
This system is a high-grade technological product. To ensure opti-
mum performance and safe working order, we recommend having it
checked by our service representative as part of regular scheduled
maintenance.
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the system
stating it is out of order and contact our service representative.
Malfunctions in the surgical microscope
T
Problem Possible cause Remedy See
No function at all. Power plug of suspension
system not inserted.
Plug in the power cord. --
Power switch of suspension
system not switched on.
Press power switch.
Green indicator light in
power switch must be on.
page 108
Automatic circuit breaker in
power switch of suspension
system has been activated.
Press power switch again. --
Line power failure. Contact in-house electri-
cian.
--
Surgical field illumination on
microscope not working.
Light guide not properly in-
serted in microscope.
Insert light guide as far as it
will go.
page 154
Failure of suspension
system electronics.
Illuminate surgical field
using an OR illuminator.
Contact service dept.
--
Lamp module in suspension
system has no contact.
Insert lamp module as far as
it will go.
page 232