What to do in the event of malfunctions 215
Failure of the zoom function
Failure of the zoom function
Motorized magnification adjustment is not possible.
• Press the Manual button.
• Use zoom adjustment knob (1) of the microscope to manually set the
magnification required (if necessary, use a tool, e.g. a screwdriver,
coin, etc.)
Malfunction of the zoom function
The zoom function performs uncontrollable settings.
• Disconnect the surgical microscope from the suspension system. The
connector is located under the cover of the suspension arm. The illu-
mination remains on.
1