Technical data
214 What to do in the event of malfunctions
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Failure of magnetic brakes
If the magnetic brakes fail (magnetic brakes are locked), you can manu-
ally position the articulated arm including the microscope by overcoming
the locking effect of the magnetic brakes.
Failure of the X-Y coupling
Failure of the X-Y coupling
Motorized control of the X-Y coupling is not possible.
• Press the Manual button.
• Manually position the surgical microscope using the movement op-
tions provided by the suspension system.
Malfunction of the X-Y coupling
The X-Y coupling performs uncontrollable movements.
• Disconnect the surgical microscope from the suspension system. The
connector is located under the cover of the suspension arm. The light
source remains on.