208 What to do in the event of malfunctions
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Lamp failure in the Superlux Eye light source
Warning!
Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
Before opening the lamp housing, make sure that the system is moved
to a position where neither the patient nor the user is put at risk by fall-
ing items.
Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronics mod-
ules. Inform our service department.
Caution:
Do not cover the ventilation grid! For example, drapes could be covering
the grid. This can lead to overheating of the lamp module and to lamp
failure.
Note:
Yellow indicator lamp (5) lights when the lamp has failed, or if the lamp
module is defective. After activation and ignition of the backup bulb, the
yellow indicator lamp turns off again.