Technical data
206 What to do in the event of malfunctions
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
What to do in an emergency
Lamp failure of the halogen light source
Caution:
Do not cover ventilation grid (2)! For example, drapes could be covering
the grid. This can lead to overheating of the lamp modules and to lamp
failure.
Note:
The lamp housing contains a backup lamp which is automatically swung
into the illumination beam path when the first lamp fails. Open flap (3) and
yellow indicator lamp (7) indicate that the backup lamp is being used.
Manual switching to the backup lamp
Press button (4) to manually activate the backup lamp.
If the backup lamp fails:
Warning!
If you change the lamp shortly after it has failed, the lamp will still be very
hot. Wear heat-protection gloves to avoid burns!
• Turn off the suspension system at the power switch.
• Press button (5) to slightly eject lamp module (1). Pull out the lamp
module and replace the lamp, or insert the lamp module of the second
light source.
Switch the suspension system back on. Adjust the brightness of the light
source on display field (6) as required.