Technical data
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Operation 197
Roll-over procedure
Use the "Roll-over procedure" to select the required function in the default
sequence of the above table.
Every time you press the relevant button "∇" you advance clockwise.
Every time you press the relevant button "Δ" you advance counterclock-
wise.
→ XYZ-RES → FOC-MEM → XY-INV →
↑↓
← XY-RES ← AUX ← PHOTO ← SDI ←
If you have assigned FOC-MEM to key C or D of the foot control panel,
you can determine by the length of time for which you press the relevant
key of the foot control panel during operation whether a positioning run is
to be triggered (press the key for less than 2 seconds) or whether the cur-
rent position is to be saved (press the key for more than 2 seconds).
Caution:
Be extremely careful when changing these settings. You should change
settings only under your own user ID. Notify all users of any changes, or
make sure that each user only works under his own user ID.
Saving parameter settings
As soon as you have entered a parameter setting, it is saved under the
current user ID.
Explanation of the assignable functions:
Path: To return to the basic mode, press the "MODE" key in configuration
mode 1 or do not press any key for 20 seconds.
To switch to configuration mode 2, simultaneously press the
"MODE" and "STORE" keys.
XY-RES The X-Y coupling moves to its center position and the
focus adopts its initial position in the focusing range.
XYZ-RES The X-Y coupling moves to its center position, the focus
adopts its initial position in the focusing range and the
zoom system adopts a position previously saved using the
STORE key.
XY-INV Inverts the movement direction of the X-Y coupling. This
is a useful function when you are using an image inverting
system in vitreo-retinal surgery. All you have to do is press
one of the two buttons (C or D) on the foot control panel to
invert the movement direction of the X-Y coupling.