Technical data

G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Operation 193
Warning!
Software and hardware failure may increase the brightness of the Su-
perlux Eye light source, leading to retinal injury in the patient's eye.
If several successive beeps are emitted and the xenon lamp is lit when
the system is switched on, this indicates a malfunction of the Superlux
Eye light source.
Attach a sign to the system stating it is out of order and contact our
service representative.
Saving the settings
As soon as you have made a setting, the setting is accepted and saved
under the current user number.
Exception: the brightness setting cannot be saved. It is reset to level 0.5
at the next power-on of the system.