Technical data

G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Operation 189
Lamp 1: xenon
Lamp 2: halogen
Acoustic signals
Service display
In the event of an error, e.g. during the software check after power-on of
the suspension system, an error message appears in the control and dis-
play panel: a wrench symbol and an error code (XXX) are displayed and
three successive beeps are heard.
Inform the service department, telling them the error code and the serial
number of the system.
Motorized control of the surgical microscope is then no longer possible.
All microscope functions except recentering of the X-Y coupling can only
be operated manually.
However, you can still release the magnetic brakes using the appropriate
button on the left and right handgrips of the microscope.
When you press the "MANUAL" button, the surgical microscope can no
longer be operated via the foot control panel, the handgrips or the control
and display panel.
You can continue using the light source, but the lamp brightness is auto-
matically adjusted to a fixed setting. You can use the foot control panel to
switch the light source on and off.
Three succes-
sive beeps
Error message during the software check after
power-on of the suspension system.
Error message for an internal system error.
One beep When the focus or zoom position is saved.
One beep After power-on of the suspension system.
Intermittent
beep
Light source failure.
MODE
STORE MANUAL
XXX