Technical data
182 Operation
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
• Replace the xenon lamp in good time.
• Reset the service hour counter to "0" after replacing the lamp.
• For the lamp change procedure and how to reset the service hour
counter, see "Switching to the backup lamp" on page 38.
Balance setting
– Check that the suspension arm has been properly balanced. When
the release button on the surgical microscope is pressed, the effort re-
quired to move the arm up or down must be the same.
Limitation of downward travel
– The minimum working distance (height) from the surgical field has
been set using the adjustment screw for limiting downward travel.
S88 floor stand only: Stand base
– Check that the brake tab has been pressed and that the stand is se-
curely locked in position.
Check the accessories
– Using the manuals provided, check that the other equipment (surgical
microscope, coobservation tube, video system, etc.) is functioning
properly.
Checking the foot control panel
– Check that the plug of the foot control panel has been connected.
– Check that the power switch of suspension system has been switched
on.
• Check that all functions assigned to the respective buttons on the foot
control panel are working properly.