Technical data
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Operation 181
– Check that image quality is the same throughout the entire magnifica-
tion range.
Check the accessories
Using the manuals provided, check that the other equipment (illumination
system, video system, etc.) is functioning properly.
Checking the suspension system
Note:
After switching on, the suspension system automatically performs a self-
test which takes approx. five seconds.
The suspension system is equipped either with a halogen light source, a
Superlux Eye light source or with a Superlux Eye light source with an ad-
ditional, integrated halogen light source (option).
Lamp brightness (halogen, xenon)
• Check that the lamp brightness display shows the minimum level (0.5)
after power-on of the system.
• Change the lamp brightness across the entire control range, and
check that brightness variation has an effect on the surgical field illu-
mination (bright/dark).
Halogen light source
– The halogen light source is on and the green indicator lamp is lit.
– The halogen lamps including the backup lamps are intact, i.e. the yel-
low indicator lamp is not lit.
Superlux Eye light source
– A beep sounds after power-on of the system, and stops when the xe-
non lamp has ignited correctly and if no other error has occurred.
– If the beep does not stop, the system must not be used.
– The xenon lamp is on and the green indicator lamp is lit.
Note
:
If the first lamp has failed and the backup lamp is in use (red segment
in the switching knob lights up), make sure to have a backup lamp
module ready at hand as a precaution.