Technical data

G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Operation 181
Check that image quality is the same throughout the entire magnifica-
tion range.
Check the accessories
Using the manuals provided, check that the other equipment (illumination
system, video system, etc.) is functioning properly.
Checking the suspension system
Note:
After switching on, the suspension system automatically performs a self-
test which takes approx. five seconds.
The suspension system is equipped either with a halogen light source, a
Superlux Eye light source or with a Superlux Eye light source with an ad-
ditional, integrated halogen light source (option).
Lamp brightness (halogen, xenon)
Check that the lamp brightness display shows the minimum level (0.5)
after power-on of the system.
Change the lamp brightness across the entire control range, and
check that brightness variation has an effect on the surgical field illu-
mination (bright/dark).
Halogen light source
The halogen light source is on and the green indicator lamp is lit.
The halogen lamps including the backup lamps are intact, i.e. the yel-
low indicator lamp is not lit.
Superlux Eye light source
A beep sounds after power-on of the system, and stops when the xe-
non lamp has ignited correctly and if no other error has occurred.
If the beep does not stop, the system must not be used.
The xenon lamp is on and the green indicator lamp is lit.
Note
:
If the first lamp has failed and the backup lamp is in use (red segment
in the switching knob lights up), make sure to have a backup lamp
module ready at hand as a precaution.