Checklist
Warning!
If a function fails, you must not use this instrument for safety reasons. Cor-
rect the fault (see the "Troubleshooting table“) or contact our service dept.
Always check the function of the system before surgery (without patient!)
using the following checklist:
Requirements:
– Check that the correct rated voltage has been set on the suspension
system.
– Check that all cables have been connected.
– Check that the light guide has been connected.
– Check that the protective cover has been removed from the micro-
scope's objective lens
• Turn on the system at the power switch of the suspension system.
Checking the surgical microscope
Check the zoom function
– Press the appropriate button on the foot control panel.
Check the focusing function
– Press the appropriate button on the foot control panel.
Friction adjustment of the surgical microscope
– Check that the friction of the surgical microscope's rotation has been
adjusted as required using the friction adjustment knob on the suspen-
sion system.
Speeds of the microscope functions
– Check that the speeds of the microscope functions have been adjust-
ed as required on the suspension system.
Eyepieces / binocular tube
– Check that the surgical microscope and the tube are in a position con-
venient for you.
– Check that the correct interpupillary distance has been set.
– Check that the eyecups have been adjusted in such a way that you
can see the full field of view.
– Check that the correct prescription has been set on the diopter scale.