Technical data

176 Preparations
G-30-1682-en OPMI® Lumera® T Issue 4.0
Printed on 02. 02. 2009
Preparing the system for sterile use
Asepsis sets
For sterile use, the system can be equipped with resterilizable products.
The asepsis sets available from Carl Zeiss contain caps and handgrips
which can be sterilized in autoclaves. For detailed information on sterili-
zation please see the enclosed instructions "Preparation of resterilizable
products".
Warning!
The products contained in the asepsis sets must be cleaned, disinfected
and sterilized before each use. This also applies to the first use after de-
livery.
The following asepsis caps can be fitted on the controls of the surgical mi-
croscope and tube:
1 Sterilizable cap
for the PD knob (adjustment of the interpupillary distance)
2 Sterilizable cap
for handgrips
3 Sterilizable cap
for control knobs with 22 mm diameter
Drapes
Sterile single-use drapes are available to cover the system.
When draping the system, make sure there is enough slack in the drapes
to allow for movement of the microscope carrier and surgical microscope.
It is especially important that the drapes are completely loose around the
handgrips. The surgeon must be able to operate the controls through the
drape.